Yes, otherwise an incorrect reading will be shown. For the Velocity Table Click Here! These are also listed in the manual supplied with the gauge.
If only a few measurements are to be taken on a material other than Steel, it may be easier to leave the calibration set for Steel and convert the readings by multiplying by the Conversion Factor for the material being measured:
True Reading = Reading x Conversion Factor
Conversion factors are listed in the Velocity Table. This method avoids unnecessary re-calibration.
|
|